Technical data sheet
Operating light evoLight 800 / 600
Table of contents:
Description.................................................................................................................. 2
Technical data ............................................................................................................ 3
Technical drawing ....................................................................................................... 5
Contact:
evonos GmbH & Co KG
Stockacher Str. 134
78532 Tuttlingen
Tel:
Fax:
+49 7461-965774-0
+49 7461-965774-89
Email:[email protected]
Confidential information - Subject to technical changes; tolerance ± 10%
V1.0
page 1 from 5
Technical data sheet
Operating light evoLight 800 / 600
Description
High-quality LEDs paired with specially dimensioned lens technology form the basis for ideal
illumination of the surgical field. The LED technology consists exclusively of visible light
without any infrared component, which makes complex filtering unnecessary. The perfect
arrangement of the LED lenses results in shadow-free illumination with simultaneous deep
light for deep body cavities.
The multicolor lights also offer a wide range of color temperature adjustment between 3,500
up to 5,500 Kelvin and achieve a high color rendering index of Ra > 96 in various settings.
The evonos operating lights are made of high-quality materials such as aluminum and ESG
safety glass. The aluminum housing favors the well-designed heat management, which is
important in two respects. Firstly, it prevents heat build-up on the LED thanks to efficient heat
dissipation, which results in an extremely long LED service life. The good heat management
and low power consumption also produce very good results in combination with laminar flow.
In addition, the evoLight surgical lights have a hygiene-optimized design, which is supported
by the seamlessly enclosed construction and the use of ESG safety glass.
Last but not least, the ESG safety glass offers the advantage of consistently high light output
over the entire service life of the LED light.
All operating functions can be carried out via the cardanic control panel:
-
On/O button
MODE button (memory function)
AMBI button (backlight for endoscopic procedures)
Illuminance plus/minus button (8 steps)
Larger/smaller light field button (infinitely variable)
Color temperature plus/minus button (5 steps)
In addition, the reusable handle can be used to perform two operating functions in parallel and
directly on the sterile handle. The assignment of the reusable handle is set to intensity and
field adjustment as standard.
Confidential information - Subject to technical changes; tolerance ± 10%
V1.0
page 2 from 5
Technical data sheet
Operating light evoLight 800 / 600
Technical data
Photometric data:
evoLight 800
evoLight 600
Central illuminance at a distance of 1m
[lx]
160.000
160.000
Electrical dimmability from/to
[lx]
Light field diameter d10 at a distance of 1m
[mm]
40.000-160.000
25-100%
170-300
40.000-160.000
25-100%
170-280
Light field diameter d50/d10 at a distance of
1m
[%]
0,5-0,6
0,5-0,6
Mechanical field adjustment
[mm]
Yes, stepless
Yes, stepless
Color temperature
[K]
Color rendering index Ra (1-8)
3.500/4.000/4.500/
5.000/5.500
96
3.500/4.000/4.500/
5.000/5.500
96
Red rendering index R9
96
96
Skin rendering index R13
98
98
Total irradiance
[W/m²]
520
510
Photometric radiation equivalent
[lm/W]
308
314
Illumination depth L1+L2 at 20%
[mm]
900
1.000
Illumination depth L1+L2 at 60%
[mm]
430
520
Working area from/to
[mm]
700-1.600
700-1.700
Laminar Flow Index (turbulence
measurement)
Temperature increase in the head area
[%]
28
24
[°C]
<1
<1
Temperature increase in the operating field
[°C]
< 10
< 10
LED light sources
[Piece]
78
54
Radiant energy
3,4
3,4
Ambient light MIC interventions
[mW/
m²lx]
[lx]
8.000
8.000
Residual illuminance with a shutter
[%]
60
56
Residual illuminance with two shades
[%]
52
48
Residual illuminance with a tube
[%]
100
100
[%]
60
56
[%]
52
48
[lx/W]
4.000
5.000
Residual illuminance with one tube and one
shutter
Residual illuminance with one tube and two
shades
Efficiency from max. light intensity to max.
power
Confidential information - Subject to technical changes; tolerance ± 10%
V1.0
page 3 from 5
Technical data sheet
Operating light evoLight 800 / 600
Electrical data:
Power supply incl. mounting plate LxWxH
evoLight 800
evoLight 600
130 x 100 x 50
220 x 145 x 50
90 - 263
Power supply Primary voltage AC
[V]
130 x 100 x 50
220 x 145 x 50
90 - 263
Input voltage at the luminaire body DC
[V]
20-32
20-32
Maximum current consumption
[A]
2,5
2,5
Typical power consumption
[A]
1,5
1,1
Power consumption at 24 V
[W]
40
32
Service life of the light sources
[h]
> 60.000
> 60.000
Classification according to MPG
I
I
Protection class according to IEC 60601
I
I
Safety class support arm system
IP 30
IP 30
Safety class Luminaire body
IP 54
IP 54
Conformity
CE
CE
Mechanical data:
evoLight 800
evoLight 600
[mm]
Minimum room height, fully cardanic (power
supply on ceiling pipe / headroom 2.25 m)
[mm]
2.700
2.700
Range of action
[mm]
1.730
1.730
Height adjustment spring arm
[mm]
1.178
1.178
approx. single weight standard DR
[kg]
10
10
Approx. single weight boom
[kg]
9,5
9,5
approx. single weight of spring arm
[kg]
7,5
7,5
approx. single weight Power Supply
[kg]
1,2
1,2
Approx. individual weight of luminaire body
[kg]
20
15
Total weight
[kg]
47,2
42,2
Vertical weight force
[N]
approx. 463
approx. 413
Torsional / bending moment
[Nm]
approx. 365
approx. 300
Canopy dimensions
[mm]
Ø560x100
Ø560x100
Height of luminaire body without sterile
handle
[mm]
approx. 55
approx. 55
Dimensions luminaire body
[mm]
Ø690
Ø585
Confidential information - Subject to technical changes; tolerance ± 10%
V1.0
page 4 from 5
Technical data sheet
Operating light evoLight 800 / 600
Technical drawing
Confidential information - Subject to technical changes; tolerance ± 10%
V1.0
page 5 from 5