Refrigerator ZL-2 TIANJIN GUOMING MEDICINAL EQUIPMENT CO.,LTD. Introduction Brief introduction ZL-2 consists of main box, compressor, condenser, cooling bath (evaporator) and fan etc. The Zl-2 refrigerator and the HW-2 constant temperature bath constitute a constant low temperature bath, which is high efficiency. It cools water from 30 ℃ to 10℃ in 45 minutes. ZL-2 main box consists of bottom, foot, box cover, front panel, and rear panel etc. Power on switch Rear panel AC input socket Name plate Water outlet ● ● ● ● ● ● ● Water inlet Front panel includes: Product model Manufacturing Rear panel includes: Power on switch AC input socket Water inlet Water outlet Name plate Function description ● Power on switch Power on switch is the main power switch. In ON position, unit is in working status, inversely, unit stops working. Before connecting to AC power, please ensure it’s in OFF position! ● AC input socket ZL-2 Refrigerator Manual 1 AC input socket is connected to the power supply with power cord offered in this unit. Please make sure AC power supply is connected properly! ● Water inlet The cooling fluids flow in the ZL-2 through this water inlet. ● Water outlet The cooling fluids flow out the ZL-2 through this water outlet. Specification Continuous work: Power: Dimension: Weight: >24hr 220V/60Hz/600W 480×380×280MM3 27KG Environment requirement Usage temperature Store temperature Humidity +18℃~+28℃ -10℃~+50℃ 20%~80% Power supply Range of input power supply Max input power 220V±10%VAC 60Hz 600W Installation Unpacking and Checking Open the package, take out manual, instrument, and accessories. And check items according packing list. Any mismatch/damage happened, please contact marketing department immediately. Unit settle It must be placed on stable work station/table and put damping pad between table and unit. Minimum space required around test is 0.5m, to ensure heat escaping. Clean Instrument periodically to prevent it from abnormal cooling. ZL-2 Refrigerator Manual 2 Pipe line connect Connect the ZL-2 and the HW-2 with attached plastic pipe. The inlet and the outlet are on the rear panel. The inlet of the ZL-2 is connected to the outlet of the HW-2, while the outlet of the ZL- 2 is connected to the inlet of the HW-2. The pipe suit with maximum temperature thimble is the shortest practical length. Soak the pipe in hot water to soft it before connecting the pipe. Connect to the power supply Link the socket & AC output socket of the HW-2 with attached power cord. Cooling fluids Suggest using distillation water over +8℃. Suggest using mixture of tap water and laboratory grade ethylene glycol below +8℃. Do not use automobile anti-freeze. Cooling bath The blocked cooling system consists of cooling bath inside ZL-2 and constant temperature bath. It gets up preset temperature in 30 minutes when distillation water is used. Operation Not open when room temperature below +5℃ . Not open without cooling fluids or no circulation. Not open when fluids temperature below +5℃ or over40℃. ZL-2 Refrigerator Manual 3 Power on Turn the power switch to I position, the power indicator lights, the bath runs. After the fluids without bubble flow smoothly, Turn the power switch of refrigerator to I position, the power indicator lights, the refrigerator runs. Power off First turn the power’s switch of the refrigerator to O position, the power indicator off, turn it off. Turn the power switch of the bath to O position, the power indicator off, the bath off. When the refrigerator shut off, allow about five minutes before restarting the unit. Otherwise it will be damaged. Maintenance Must clean box and bath periodically, do not use steel brush, it result in damage to the equipment. Clean it at least on a month. ZL-2 Refrigerator Manual 4