8 Liquichek Cardiac Markers Plus Control LT Levels 1, 2, 3, 1A, 1B and 1C TM 146 147 148 27104 27105 27106 Level 1 Level 2 Level 3 Level 1A Level 1B Level 1C 6 x 3 mL 6 x 3 mL 6 x 3 mL 6 x 3 mL 6 x 3 mL 6 x 3 mL 146X 147X 148X 27104X 27105X 27106X Level 1 MiniPak Level 2 MiniPak Level 3 MiniPak Level 1A MiniPak Level 1B MiniPak Level 1C MiniPak 1 x 3 mL 1 x 3 mL 1 x 3 mL 1 x 3 mL 1 x 3 mL 1 x 3 mL 23660 2020-01-31 Level 1 Level 2 Level 3 Level 1A Level 1B Level 1C 23661 23662 23663 23664 23665 23666 http://www.myeinserts.com/23660 MÉTODO (1) Nivel 1 - 23661 Unidades Media Rango ISOENZIMA CK-MB (ACTIVIDAD) (2) Siemens ADVIA Chemistry Systems (DiaSys) U/L 8.35 <5.00 – 13.1 Siemens Dimension Series (MBI) U/L 8.51 5.78 – 11.2 Siemens ADVIA Chemistry Systems (wrCRP) mg/dL 0.049 0.032 – 0.067 Siemens Dimension Series (RCRP) mg/dL 0.129 0.073 – 0.185 ng/mL 0.081 0.048 – 0.114 PROTEÍNA C REACTIVA (PCR) TROPONINA I Siemens ADVIA Centaur CP (TnI Ultra) (Rgt. Lots 010110 and above) MÉTODO (1) Nivel 2 - 23662 Unidades Media Rango ISOENZIMA CK-MB (ACTIVIDAD) (2) Siemens ADVIA Chemistry Systems (DiaSys) U/L 23.9 17.0 – 30.8 Siemens Dimension Series (MBI) U/L 26.3 22.1 – 30.5 Siemens ADVIA Chemistry Systems (wrCRP) mg/dL 0.203 0.168 – 0.237 Siemens Dimension Series (RCRP) mg/dL 0.249 0.196 – 0.303 ng/mL 2.82 2.17 – 3.46 PROTEÍNA C REACTIVA (PCR) TROPONINA I Siemens ADVIA Centaur CP (TnI Ultra) (Rgt. Lots 010110 and above) MÉTODO (1) Nivel 3 - 23663 Unidades Media Rango ISOENZIMA CK-MB (ACTIVIDAD) (2) Siemens ADVIA Chemistry Systems (DiaSys) U/L 73.0 57.3 – 88.6 Siemens Dimension Series (MBI) U/L 82.2 73.7 – 90.8 Siemens ADVIA Chemistry Systems (wrCRP) mg/dL 0.804 0.705 – 0.903 Siemens Dimension Series (RCRP) mg/dL 0.717 0.620 – 0.813 ng/mL 17.9 14.0 – 21.8 PROTEÍNA C REACTIVA (PCR) TROPONINA I Siemens ADVIA Centaur CP (TnI Ultra) (Rgt. Lots 010110 and above) Página 1 ESPAÑOL/Spanish • 2017-04 • 16000202-00 8 Liquichek Cardiac Markers Plus Control LT, Levels 1, 2, 3, 1A, 1B and 1C TM MÉTODO (1) Nivel 1A - 23664 Unidades Media Rango ISOENZIMA CK-MB (ACTIVIDAD) (2) Siemens ADVIA Chemistry Systems (DiaSys) U/L 8.26 <5.00 – 13.0 Siemens Dimension Series (MBI) U/L 8.69 5.96 – 11.4 Siemens ADVIA Chemistry Systems (wrCRP) mg/dL 0.050 0.031 – 0.068 Siemens Dimension Series (RCRP) mg/dL 0.160 0.122 – 0.198 ng/mL 0.229 0.174 – 0.284 PROTEÍNA C REACTIVA (PCR) TROPONINA I Siemens ADVIA Centaur CP (TnI Ultra) (Rgt. Lots 010110 and above) MÉTODO (1) Nivel 1B - 23665 Unidades Media Rango ISOENZIMA CK-MB (ACTIVIDAD) (2) Siemens ADVIA Chemistry Systems (DiaSys) U/L 8.59 <5.00 – 13.5 Siemens Dimension Series (MBI) U/L 8.73 5.99 – 11.5 Siemens ADVIA Chemistry Systems (wrCRP) mg/dL 0.049 0.031 – 0.067 Siemens Dimension Series (RCRP) mg/dL 0.135 0.074 – 0.197 ng/mL 0.024 <0.006 – 0.047 PROTEÍNA C REACTIVA (PCR) TROPONINA I Siemens ADVIA Centaur CP (TnI Ultra) (Rgt. Lots 010110 and above) MÉTODO (1) Nivel 1C - 23666 Unidades Media Rango ISOENZIMA CK-MB (ACTIVIDAD) (2) Siemens ADVIA Chemistry Systems (DiaSys) U/L 8.07 <5.00 – 12.8 Siemens Dimension Series (MBI) U/L 8.38 5.67 – 11.1 Siemens ADVIA Chemistry Systems (wrCRP) mg/dL 0.048 0.030 – 0.065 Siemens Dimension Series (RCRP) mg/dL 0.132 0.078 – 0.186 ng/mL 0.033 0.008 – 0.058 PROTEÍNA C REACTIVA (PCR) TROPONINA I Siemens ADVIA Centaur CP (TnI Ultra) (Rgt. Lots 010110 and above) NOTAS AL PIE DE PÁGINA (1) Es posible que no todas las notas al pie sean de aplicación para la tabla de datos personalizada que ha seleccionado. (2) Los valores se obtuvieron a 37 °C. (3) Los valores asignados se determinaron utilizando el protocolo del fabricante del reactivo o instrumento y puede que no representen rangos de ±3s. (4) El fabricante del kit de reactivos solicita que se consulten las FICHAS DE DATOS DE CONTROL del kit de reactivos de LSI MEDIENCE PATHFAST para la asignación de valores. (5) Este producto debe ensayarse con dilución automática, igual que las muestras de pacientes. (6) Debe efectuarse una dilución antes de llevar a cabo las pruebas cuando el valor indicado excede el estándar más alto. (7) La recuperación del analito puede quedar fuera del rango analítico del ensayo. En este momento no se dispone de datos. Consulte cualquier duda. § Debido a la limitada participación, no se obtuvieron los datos necesarios para calcular las medias y los rangos aceptables para este ensayo. Si su centro está interesado en participar en el Programa de Asignación de Valores de este ensayo, pónganse en contacto con su oficina local de Bio-Rad. No aplicable Página 2 ESPAÑOL/Spanish • 2017-04 • 16000202-00