Liquichek Cardiac Markers Plus Control LT
Levels 1, 2, 3, 1A, 1B and 1C
TM
146
147
148
27104
27105
27106
Level 1
Level 2
Level 3
Level 1A
Level 1B
Level 1C
6 x 3 mL
6 x 3 mL
6 x 3 mL
6 x 3 mL
6 x 3 mL
6 x 3 mL
146X
147X
148X
27104X
27105X
27106X
Level 1 MiniPak
Level 2 MiniPak
Level 3 MiniPak
Level 1A MiniPak
Level 1B MiniPak
Level 1C MiniPak
1 x 3 mL
1 x 3 mL
1 x 3 mL
1 x 3 mL
1 x 3 mL
1 x 3 mL
23660
2020-01-31
Level 1
Level 2
Level 3
Level 1A
Level 1B
Level 1C
23661
23662
23663
23664
23665
23666
http://www.myeinserts.com/23660
MÉTODO (1)
Nivel 1 - 23661
Unidades
Media
Rango
ISOENZIMA CK-MB (ACTIVIDAD) (2)
Siemens ADVIA Chemistry Systems (DiaSys)
U/L
8.35
<5.00
–
13.1
Siemens Dimension Series (MBI)
U/L
8.51
5.78
–
11.2
Siemens ADVIA Chemistry Systems (wrCRP)
mg/dL
0.049
0.032
–
0.067
Siemens Dimension Series (RCRP)
mg/dL
0.129
0.073
–
0.185
ng/mL
0.081
0.048
–
0.114
PROTEÍNA C REACTIVA (PCR)
TROPONINA I
Siemens ADVIA Centaur CP (TnI Ultra) (Rgt. Lots 010110 and above)
MÉTODO (1)
Nivel 2 - 23662
Unidades
Media
Rango
ISOENZIMA CK-MB (ACTIVIDAD) (2)
Siemens ADVIA Chemistry Systems (DiaSys)
U/L
23.9
17.0
–
30.8
Siemens Dimension Series (MBI)
U/L
26.3
22.1
–
30.5
Siemens ADVIA Chemistry Systems (wrCRP)
mg/dL
0.203
0.168
–
0.237
Siemens Dimension Series (RCRP)
mg/dL
0.249
0.196
–
0.303
ng/mL
2.82
2.17
–
3.46
PROTEÍNA C REACTIVA (PCR)
TROPONINA I
Siemens ADVIA Centaur CP (TnI Ultra) (Rgt. Lots 010110 and above)
MÉTODO (1)
Nivel 3 - 23663
Unidades
Media
Rango
ISOENZIMA CK-MB (ACTIVIDAD) (2)
Siemens ADVIA Chemistry Systems (DiaSys)
U/L
73.0
57.3
–
88.6
Siemens Dimension Series (MBI)
U/L
82.2
73.7
–
90.8
Siemens ADVIA Chemistry Systems (wrCRP)
mg/dL
0.804
0.705
–
0.903
Siemens Dimension Series (RCRP)
mg/dL
0.717
0.620
–
0.813
ng/mL
17.9
14.0
–
21.8
PROTEÍNA C REACTIVA (PCR)
TROPONINA I
Siemens ADVIA Centaur CP (TnI Ultra) (Rgt. Lots 010110 and above)
Página 1
ESPAÑOL/Spanish • 2017-04 • 16000202-00
8
Liquichek Cardiac Markers Plus Control LT, Levels 1, 2, 3, 1A, 1B and 1C
TM
MÉTODO (1)
Nivel 1A - 23664
Unidades
Media
Rango
ISOENZIMA CK-MB (ACTIVIDAD) (2)
Siemens ADVIA Chemistry Systems (DiaSys)
U/L
8.26
<5.00
–
13.0
Siemens Dimension Series (MBI)
U/L
8.69
5.96
–
11.4
Siemens ADVIA Chemistry Systems (wrCRP)
mg/dL
0.050
0.031
–
0.068
Siemens Dimension Series (RCRP)
mg/dL
0.160
0.122
–
0.198
ng/mL
0.229
0.174
–
0.284
PROTEÍNA C REACTIVA (PCR)
TROPONINA I
Siemens ADVIA Centaur CP (TnI Ultra) (Rgt. Lots 010110 and above)
MÉTODO (1)
Nivel 1B - 23665
Unidades
Media
Rango
ISOENZIMA CK-MB (ACTIVIDAD) (2)
Siemens ADVIA Chemistry Systems (DiaSys)
U/L
8.59
<5.00
–
13.5
Siemens Dimension Series (MBI)
U/L
8.73
5.99
–
11.5
Siemens ADVIA Chemistry Systems (wrCRP)
mg/dL
0.049
0.031
–
0.067
Siemens Dimension Series (RCRP)
mg/dL
0.135
0.074
–
0.197
ng/mL
0.024
<0.006
–
0.047
PROTEÍNA C REACTIVA (PCR)
TROPONINA I
Siemens ADVIA Centaur CP (TnI Ultra) (Rgt. Lots 010110 and above)
MÉTODO (1)
Nivel 1C - 23666
Unidades
Media
Rango
ISOENZIMA CK-MB (ACTIVIDAD) (2)
Siemens ADVIA Chemistry Systems (DiaSys)
U/L
8.07
<5.00
–
12.8
Siemens Dimension Series (MBI)
U/L
8.38
5.67
–
11.1
Siemens ADVIA Chemistry Systems (wrCRP)
mg/dL
0.048
0.030
–
0.065
Siemens Dimension Series (RCRP)
mg/dL
0.132
0.078
–
0.186
ng/mL
0.033
0.008
–
0.058
PROTEÍNA C REACTIVA (PCR)
TROPONINA I
Siemens ADVIA Centaur CP (TnI Ultra) (Rgt. Lots 010110 and above)
NOTAS AL PIE DE PÁGINA
(1) Es posible que no todas las notas al pie sean de aplicación para la tabla de datos personalizada que ha seleccionado.
(2) Los valores se obtuvieron a 37 °C.
(3) Los valores asignados se determinaron utilizando el protocolo del fabricante del reactivo o instrumento y puede que no representen rangos de ±3s.
(4) El fabricante del kit de reactivos solicita que se consulten las FICHAS DE DATOS DE CONTROL del kit de reactivos de LSI MEDIENCE PATHFAST para la
asignación de valores.
(5) Este producto debe ensayarse con dilución automática, igual que las muestras de pacientes.
(6) Debe efectuarse una dilución antes de llevar a cabo las pruebas cuando el valor indicado excede el estándar más alto.
(7) La recuperación del analito puede quedar fuera del rango analítico del ensayo.
En este momento no se dispone de datos. Consulte cualquier duda.
§ Debido a la limitada participación, no se obtuvieron los datos necesarios para calcular las medias y los rangos aceptables para este ensayo. Si su centro está
interesado en participar en el Programa de Asignación de Valores de este ensayo, pónganse en contacto con su oficina local de Bio-Rad.
No aplicable
Página 2
ESPAÑOL/Spanish • 2017-04 • 16000202-00