Programas de control de calidad en imágenes digitales.

Anuncio
Tests Operacionales
en Radiografía
Mauricio Vergara
Clínica del Country, 2015
Funcionamiento
Circuito del
filamento Negativo
Reloj seg
Ampolla de vidrio
(catodo)
Electrones
acelerados
100 mA
200 mA
300 mA
Circuito de
Alta tensión
Positivo
(anodo)
Tungsteno
FOTONES
Calor
electrones
Energía: circuito kV
FOTONES
N°: circuito mAs
110 kV, 5 mAs
22
cm
110 kV, 5 mAs
28
cm
110 kV, 12 mAs
28
cm
Control de Exposición Automático
Paciente
Tubo RX
fotones
CI
Detector
IMAGEN
fotones
tiempo mAs
AEC
N° fotones
emergentes
(ajuste densidad)
Tests Operacionales en RX
• Tests Operacional: certificar que el
equipo opera “normalmente” y
genera imágenes en rangos de
calidad clínica
• Test Aceptación: certificar que las
especificaciones técnicas actuales
del equipo son compatibles con las
originales
VISUAL
Colimador
Grilla
Grilla
AEC
Detector
R
X
gain
kV
generador
mAs
DOSIMETRO
kV
mAs
Gain
kV
Dosis : mGy, Gy, nGy
Tiempo: sec, msec
Generador kVp
Tubo RX
dosímetro
Separador
10 cm
mesa
nomina
l
50
kVp
80
kVp
110
medido
difercia
%
Filamento mAs vs Dosis
Tubo RX
dosímetro
Separador
10 cm
mesa
mAs
80
kVp
6
80
kVp
20
80
kVp
30
Gy
Colimación
Tubo RX
50 kV, 6 mAs
campo Luz
15 cm x 15 cm
detector
delimitador
mesa
Campo radiación
16 cm x 16 cm máximo
Colimación
Alineacion
Tubo RX - Detecto
mesa
detector
Histograma Fantoma
Columna Lumbar
81 kV 1.87 mAs
OBSERVACIONES
• kVp y mAs con registros OK. Equipo RX OK
• Ajuste de Densidad con valores asimétricos
• Ajuste ganancias AEC: valores muy altos y
diferentes en mesa y mural (detector
central)
• Histograma con forma OK. mAs muy bajo.
Imagen OK
• No hay antecedentes de registros de kVp,
mAs, AEC e Histogramas para comparar
• No hay especificaciones de grillas y detector
CIRS mammography phantom
TEST OPERACIONAL - CASSETTES RAYOS X
Institución: Clinica del Country
Fecha:
Equipo
Modelo
Sede Fabricante
CR
CAPSULA X
CLINICA
Fujifilm
cassette
14x17
TAMAÑO (IP 35x43 cm) Fantoma mm =180
nominal
medido
ajuste
mm
180 180 0
180
VALOR S para 1 mR
1 Sv = 100 mR
exposición kVp 80
mA 100
mAs 2
dosis mR
1
1.014
FCR S Value Control Table
2
1.026
average
1.0324
3
1.057
4
1.038
5
1.027
Exponer IP 35x43 cm (leer monitor)
S nominal 200 S min
154
S medido 279 S max
260
S corregido
200
Dic,2012
gamma is the relationship between the brightness of a pixel as it
appears on the screen, and the numerical value of that pixel.
You probably already know that a pixel can have any 'value' of Red,
Green, and Blue between 0 and 255, and you would therefore think
that a pixel value of 127 would appear as half of the maximum
possible brightness, and that a value of 64 would represent onequarter brightness, and so on. Well, that's just not the case,
Cathode-ray tubes, such as the screen you're probably reading this on
at the moment, have a peculiar relationship between the voltage
applied to them, and the amount of light emitted. It isn't linear, and in
fact it follows what's called by mathematicians and other geeks, a
'power law' (a number raised to a power). The numerical value of that
power is what we call the gamma of the monitor or system.
Imagen Matriz
impresora
numeros
Señales
electricas
DAC
Señales
electricas
monitor
Impresora
 curve
 curve
Procesador
Señales
electricas
DAC
Señales
electricas
LUT
 curve
Monitor
 curve
LUT
Resultados de curvas Gamma para sistema
radiografía digital tipo CR. La recta es la
Compuesta ideal; la curva roja es la
Compuesta medida.
TEST OPERACIONAL
Institución
:
ONCOLOGICO
Fecha
:
7/19/2012
Modelo
Equipo
Fabricante
impresora DryPix
FUJIFILM
8 x 10
nominal
mm
medido
mm
diferencia
TAMAÑO
A
156.00
155.0
1%
B
202.00
201.0
0%
DENSIDADES REFERENCIA
referencia OD
OD medida diferencia OD interno
0
0.20
0.19
5%
0.17
1
0.48
0.53
-10%
0.48
2
1.20
1.27
-6%
1.19
CONCLUSIONES
• Tests Operacionales se ejecutan en
60 minutos y requieren dosímetro
(Unfors) y fantomas (CIRS en
mamografía y fantoma columna de
densitómetro)
• Entregan información de Generador
(kVp y mAs) y calidad de imagen
• Frecuencia cada 4 meses
PROYECTO EN CURSO
HISTOGRAMAS EN
ECOGRAFIA
Performance Measurements:
Dead Zone
Horizontal Distance
Vertical Distance
Depth of Penetration
Image Uniformity
Axial Resolution
Lateral Resolution
Anechoic Mass Resolution
Gray Scale Contrast Resolution
Elasticity Image Evaluation
Transductor convexo
Transductor OK
Con postprocesamiento
Transductor OK
Sin postprocesamiento
Transductor convexo
Transductor OK
Sin postprocesamiento
Daño en cristales
Sin postprocesamiento
Transductor lineal
Sin postprocesamiento