Tests Operacionales en Radiografía Mauricio Vergara Clínica del Country, 2015 Funcionamiento Circuito del filamento Negativo Reloj seg Ampolla de vidrio (catodo) Electrones acelerados 100 mA 200 mA 300 mA Circuito de Alta tensión Positivo (anodo) Tungsteno FOTONES Calor electrones Energía: circuito kV FOTONES N°: circuito mAs 110 kV, 5 mAs 22 cm 110 kV, 5 mAs 28 cm 110 kV, 12 mAs 28 cm Control de Exposición Automático Paciente Tubo RX fotones CI Detector IMAGEN fotones tiempo mAs AEC N° fotones emergentes (ajuste densidad) Tests Operacionales en RX • Tests Operacional: certificar que el equipo opera “normalmente” y genera imágenes en rangos de calidad clínica • Test Aceptación: certificar que las especificaciones técnicas actuales del equipo son compatibles con las originales VISUAL Colimador Grilla Grilla AEC Detector R X gain kV generador mAs DOSIMETRO kV mAs Gain kV Dosis : mGy, Gy, nGy Tiempo: sec, msec Generador kVp Tubo RX dosímetro Separador 10 cm mesa nomina l 50 kVp 80 kVp 110 medido difercia % Filamento mAs vs Dosis Tubo RX dosímetro Separador 10 cm mesa mAs 80 kVp 6 80 kVp 20 80 kVp 30 Gy Colimación Tubo RX 50 kV, 6 mAs campo Luz 15 cm x 15 cm detector delimitador mesa Campo radiación 16 cm x 16 cm máximo Colimación Alineacion Tubo RX - Detecto mesa detector Histograma Fantoma Columna Lumbar 81 kV 1.87 mAs OBSERVACIONES • kVp y mAs con registros OK. Equipo RX OK • Ajuste de Densidad con valores asimétricos • Ajuste ganancias AEC: valores muy altos y diferentes en mesa y mural (detector central) • Histograma con forma OK. mAs muy bajo. Imagen OK • No hay antecedentes de registros de kVp, mAs, AEC e Histogramas para comparar • No hay especificaciones de grillas y detector CIRS mammography phantom TEST OPERACIONAL - CASSETTES RAYOS X Institución: Clinica del Country Fecha: Equipo Modelo Sede Fabricante CR CAPSULA X CLINICA Fujifilm cassette 14x17 TAMAÑO (IP 35x43 cm) Fantoma mm =180 nominal medido ajuste mm 180 180 0 180 VALOR S para 1 mR 1 Sv = 100 mR exposición kVp 80 mA 100 mAs 2 dosis mR 1 1.014 FCR S Value Control Table 2 1.026 average 1.0324 3 1.057 4 1.038 5 1.027 Exponer IP 35x43 cm (leer monitor) S nominal 200 S min 154 S medido 279 S max 260 S corregido 200 Dic,2012 gamma is the relationship between the brightness of a pixel as it appears on the screen, and the numerical value of that pixel. You probably already know that a pixel can have any 'value' of Red, Green, and Blue between 0 and 255, and you would therefore think that a pixel value of 127 would appear as half of the maximum possible brightness, and that a value of 64 would represent onequarter brightness, and so on. Well, that's just not the case, Cathode-ray tubes, such as the screen you're probably reading this on at the moment, have a peculiar relationship between the voltage applied to them, and the amount of light emitted. It isn't linear, and in fact it follows what's called by mathematicians and other geeks, a 'power law' (a number raised to a power). The numerical value of that power is what we call the gamma of the monitor or system. Imagen Matriz impresora numeros Señales electricas DAC Señales electricas monitor Impresora curve curve Procesador Señales electricas DAC Señales electricas LUT curve Monitor curve LUT Resultados de curvas Gamma para sistema radiografía digital tipo CR. La recta es la Compuesta ideal; la curva roja es la Compuesta medida. TEST OPERACIONAL Institución : ONCOLOGICO Fecha : 7/19/2012 Modelo Equipo Fabricante impresora DryPix FUJIFILM 8 x 10 nominal mm medido mm diferencia TAMAÑO A 156.00 155.0 1% B 202.00 201.0 0% DENSIDADES REFERENCIA referencia OD OD medida diferencia OD interno 0 0.20 0.19 5% 0.17 1 0.48 0.53 -10% 0.48 2 1.20 1.27 -6% 1.19 CONCLUSIONES • Tests Operacionales se ejecutan en 60 minutos y requieren dosímetro (Unfors) y fantomas (CIRS en mamografía y fantoma columna de densitómetro) • Entregan información de Generador (kVp y mAs) y calidad de imagen • Frecuencia cada 4 meses PROYECTO EN CURSO HISTOGRAMAS EN ECOGRAFIA Performance Measurements: Dead Zone Horizontal Distance Vertical Distance Depth of Penetration Image Uniformity Axial Resolution Lateral Resolution Anechoic Mass Resolution Gray Scale Contrast Resolution Elasticity Image Evaluation Transductor convexo Transductor OK Con postprocesamiento Transductor OK Sin postprocesamiento Transductor convexo Transductor OK Sin postprocesamiento Daño en cristales Sin postprocesamiento Transductor lineal Sin postprocesamiento